Controlling data variation

Daphne Allen

December 9, 2015

2 Min Read
Packaging Digest logo in a gray background | Packaging Digest

When testing package seals according to ASTM F2054 (Standard Test Method for Burst Testing of Flexible Package Seals Using Internal Air Pressurization Within Restraining Plates), Thomas McNabb and his team of packaging engineers at Edwards Lifesciences would often see excessive variation in the results data as well as nonnormal data and inconsistent results over time. He decided to investigate to determine whether their testing approach could be improved to reduce variation.

McNabb, who serves as a principal packaging engineer for Edwards, spoke at HealthPack 2015 during the presentation “Burst Testing: Finding Better Value in the Data.”

McNabb explained that his team uses burst testing to evaluate seal strength during IQ, OQ, PQ as well as during in-process testing. He told the audience that if correlations were to be made between tests and testing equipment, the parameters of each test must be equivalent.

First, for this particular exercise his team held all of the factors constant: package size, material, rate of air flow into the package, pressure detection sensing mechanism, sensitivity (machine response to pressure drop), and position of test article. “The only variable was gap,” he says. “We did investigate those factors in different studies, but did not find the same sensitivity or impact on variability.”

Gap settings were then evaluated to determine whether differences had an effect on mean output value, McNabb explained. Experimenting with the gap, his team found that “gap settings do affect the value of the burst output for the same package,” he said. “As the gap increases, the burst value decreases.”

Comparing the results from testing different packages, the team found that “as the gap changes, data variation will initially increase, but the variation will eventually reach a minimum,” McNabb said.

As a result, he believes that “reduced variation leads to more accurate measurements.” He advises other users of ASTM F2054 to “keep all variables constant, including the gap.”

In addition to keeping the gap constant when measuring a particular package configuration, “the recommendation is to follow the ASTM guidance to evaluate gaps for suitability,” he adds.

McNabb’s team will next try to correlate the approach to peel testing following F88 (Standard Test Method for Seal Strength of Flexible Barrier Materials). “The idea is to improve all test methods and to get a better window into controlling data variation,” he said.

 

About the Author

Daphne Allen

Design News

Daphne Allen is editor-in-chief of Design News. She previously served as editor-in-chief of MD+DI and of Pharmaceutical & Medical Packaging News and also served as an editor for Packaging Digest. Daphne has covered design, manufacturing, materials, packaging, labeling, and regulatory issues for more than 20 years. She has also presented on these topics in several webinars and conferences, most recently discussing design and engineering trends at MD&M West 2024 and leading an Industry ShopTalk discussion during the show on artificial intelligence. She will be moderating the upcoming webinar, Best Practices in Medical Device Engineering and will be leading an Automation Tour at Advanced Manufacturing Minneapolis. She will also be attending DesignCon and MD&M West 2025.

Daphne has previously participated in meetings of the IoPP Medical Device Packaging Technical Committee and served as a judge in awards programs held by The Tube Council and the Healthcare Compliance Packaging Council. She also received the Bert Moore Excellence in Journalism Award in the AIM Awards in 2012.

Follow Daphne on X at @daphneallen and reach her at [email protected].

Sign up for the Packaging Digest News & Insights newsletter.

You May Also Like